5.2 PURE TONE TESTING
This section gives a short description of how to operate the MIDIMATE 622
Diagnostic Audiometer for both air and bone conduction and free-field
threshold measurements using pure tones.
For a more comprehensive treatment of this subject, please refer to 6. PURE
TONE AUDIOMETRY.
5.2.1 AIR CONDUCTION THRESHOLD MEASUREMENTS
1. Plug the headphones, patient microphone, patient response signal and
monitor headset (or Talk-Over microphone) into their respective
connectors on the rear panel (see Figure 1 on page 4).
2. Connect the MIDIMATE 622 to the mains.
3. Switch on the instrument by pressing the power switch on the right side
panel.
4. After the power-up test has been completed, the audiometer will
automatically access User Test 1. Proceed with this preprogrammed
test or press
MENU
to access any other User Test or press
MENU
again
to access the Select Function Menu and set up your own test as
described in 4. Test Setup.
MENU
Ensure that Air has been selected as Output. Threshold testing normally
commences at 1000 Hz in the better ear and continues with the higher
frequencies.
Air Output
5. Place the patient in sound cabin, if available. Ensure that the patient is
facing away from the audiometer.
6. Instruct the patient in usage of Patient Response Handswitch and/or
microphone and fit headphones ensuring that the red phone is on the
right ear.
7. Proceed with test starting with the better ear.
8. If the difference between the two ears is greater than 40-60 dB, repeat
the test using masking. Softkey 4 toggles masking on/off.
Softkey 4
5.2.2 BONE CONDUCTION THRESHOLD MEASUREMENTS
1. Plug the headphones, patient microphone, patient response signal,
bone conductor, insert phone and monitor headset (or Talk-Over
microphone) into their respective connectors on the rear panel.
2. Connect the MIDIMATE 622 to the mains.
3. Switch on the instrument by pressing the power switch on the right side
panel.
MIDIMATE 622 OPERATION MANUAL
5. Operating in Test Mode
29
Содержание MADSEN Midimate 622
Страница 1: ...Midimate 622 Clinical Diagnostic Audiometer User Manual Doc no 7 25 9000 09 Part No 7 25 900 0459 ...
Страница 16: ...10 MIDIMATE 622 OPERATION MANUAL 2 Installing Midimate 622 ...
Страница 18: ...Fig 3 The Selection Tree MIDIMATE 622 OPERATION MANUAL 3 General Description 12 Menu Menu Menu ...
Страница 54: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 7 Speech Audiometry 48 ...
Страница 70: ...This page has been intentionally left blank MIDIMATE 622 OPERATION MANUAL 10 Power On Options 64 ...
Страница 74: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 12 Calibration 68 ...
Страница 101: ...ZODIAC 901 Connection To 74 ZODIAC 901 Printout To 59 73 MIDIMATE 622 OPERATION MANUAL Subject Index 95 ...