2. From Test Mode (any test you have set up), press Softkey 3 to access
the Options Menu and the following will be displayed (although the
options available for Softkeys 4 and 5 will vary):
SOFTKEY 3
Options Menu:
1.5s On
1 dB
Progr
HLL-On PTS-On
3. Press Softkey 3 to start programming and the following appears on the
display:
SOFTKEY 1
Programming User Test !
Continue
Quit
4. Pressing “Quit” takes you out of the programming procedure but the
audiometer is still in the Programming Mode, i.e. Progr is still an option in
the Options menu.
Quit
The programming option will be retained until the instrument is switched
off.
5. Press Softkey 1 to continue and the following menu is displayed:
SOFTKEY 1
Select Program No.:
Test1
Test2
Test3
Test4
Test5
NOTE
: This display is your last opportunity to change your mind about
overwriting a previously stored test! If you decide
NOT
to save your current
test setup instead of the existing User Test, press
MENU
.
6. Select whichever User Test you wish to overwrite and press the
corresponding softkey. The audiometer will briefly display a message
confirming that the test has been saved as Test No. 1 to 5 and then
revert to Test Mode in the User Test selected.
Select & Save
User Test
Setup saved as Test : X
7. Check each test after having saved it to ensure that all test parameters
have been set up as desired. If not, repeat the procedure described
above.
Check Test &
Overwrite, If
Necessary
8. A template for the purpose of hanging over the five softkeys is supplied
with the audiometer-write the details of each test setup saved as a User
Test next to the corresponding softkey.
User Test Template
MIDIMATE 622 OPERATION MANUAL
11. Test Setup Programming
66
Содержание MADSEN Midimate 622
Страница 1: ...Midimate 622 Clinical Diagnostic Audiometer User Manual Doc no 7 25 9000 09 Part No 7 25 900 0459 ...
Страница 16: ...10 MIDIMATE 622 OPERATION MANUAL 2 Installing Midimate 622 ...
Страница 18: ...Fig 3 The Selection Tree MIDIMATE 622 OPERATION MANUAL 3 General Description 12 Menu Menu Menu ...
Страница 54: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 7 Speech Audiometry 48 ...
Страница 70: ...This page has been intentionally left blank MIDIMATE 622 OPERATION MANUAL 10 Power On Options 64 ...
Страница 74: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 12 Calibration 68 ...
Страница 101: ...ZODIAC 901 Connection To 74 ZODIAC 901 Printout To 59 73 MIDIMATE 622 OPERATION MANUAL Subject Index 95 ...