5.3.2 SPEECH AUDIOMETRY WITH LIVE VOICE SIGNAL
With a live voice signal it is possible to perform either threshold measure-
ments or discrimination tests and to present the stimuli via Air, Bone or Free
Field. Please refer to sections 7.1 and 7.2 for details concerning threshold
measurements and discrimination testing in speech audiometry.
1. Plug the headphones, patient microphone, patient response signal,
bone conductor and monitor headset (or Talk-Over microphone) into
their respective connectors on the rear panel.
2. Connect the MIDIMATE 622 to the mains.
3. Switch on the instrument by pressing the power switch on the right side
panel.
4. After the power-up test has been completed, the audiometer will
automatically access User Test 1. Proceed with this preprogrammed
test if suitable or press
MENU
to access any other suitable User Test or
press
MENU
again to access the Select Function Menu and set up your
own test as described in 4. Test Setup.
Output may be Air, Bone or even Free Field but
Micr.
must always be
selected as input.
5. Adjust microphone level by reading from the word list while at the same
time pressing the
MICROPHONE
pushbutton and turning one of the
rotary knobs so that the stimulus level on the Stimulus LED deflects up to
O VU. See 5.1.3 on page 20 for detailed description.
MICROPHONE
6. Place the patient in sound cabin facing away from the audiometer.
7. Instruct the patient to repeat the words he hears and test each ear in turn
(using
L/R Shift
).
L/R SHIFT
Refer to section 7.3 for further details on live voice testing.
5.3.3 SPEECH AUDIOMETRY WITH PRERECORDED SIGNAL
With a prerecorded signal it is possible to perform either threshold
measurements or discrimination tests and to present the stimuli via Air, Bone
or Free Field. Please refer to sections 7.1 and 7.2 for details concerning
threshold measurements and discrimination testing in speech audiometry.
1. Plug the headphones, patient microphone, patient response signal,
bone conductor and monitor headset (or Talk-Over microphone) into
their respective connectors on the rear panel.
2. Connect the tape recorder or CD player to the socket marked Tape 1 on
the rear panel. If masking with a prerecorded signal is to be used,
connect the signal to the socket marked Tape 2.
3. Connect the MIDIMATE 622 to the mains.
MIDIMATE 622 OPERATION MANUAL
5. Operating in Test Mode
32
Содержание MADSEN Midimate 622
Страница 1: ...Midimate 622 Clinical Diagnostic Audiometer User Manual Doc no 7 25 9000 09 Part No 7 25 900 0459 ...
Страница 16: ...10 MIDIMATE 622 OPERATION MANUAL 2 Installing Midimate 622 ...
Страница 18: ...Fig 3 The Selection Tree MIDIMATE 622 OPERATION MANUAL 3 General Description 12 Menu Menu Menu ...
Страница 54: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 7 Speech Audiometry 48 ...
Страница 70: ...This page has been intentionally left blank MIDIMATE 622 OPERATION MANUAL 10 Power On Options 64 ...
Страница 74: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 12 Calibration 68 ...
Страница 101: ...ZODIAC 901 Connection To 74 ZODIAC 901 Printout To 59 73 MIDIMATE 622 OPERATION MANUAL Subject Index 95 ...