8. SPECIAL TESTS
From the Select Function Menu, select:
Special
and the following display appears:
Select special Test :
SISI
Fowler Stenger
Rainv.
Select required test by pressing the appropriate softkey.
8.1 SHORT INCREMENT SENSITIVITY INDEX (SISI) TEST
Select the following setup:
Special, SISI, Left or Right.
The following display appears:
Freq.
SISI (A) L:Tone
Score
0 ( %)
1000 Hz
1-5dB
20 dB
Pres.:
0 Stop
Softkey 2 = Intensity 1 dB/5 dB
Softkey 5 = Stop to reset or continue
The MIDIMATE 622 is able to perform automatic SISI tests with the
standardized specifications for this test.
Explain to the patient that he will hear a continuous tone the intensity of which
will be increased for a very short time (200 ms) every fifth second and that he
should press the patient signal
immediately
each time he hears such a
change. To demonstrate this, ensure that modulation is set at the default
position 5 dB where it should be easy for the patient to hear the intensity
changes and set HL at 40 dB or higher.
Press Tone Switch to start the test. The HL setting on the display will change
each time the intensity is raised. Changing Frequency or Hearing Level will
stop the SISI test-restart by pressing the Tone Switch.
TONE
The display will show the number of presentations, the patient's score and
the score as a percentage of presentations.
When the patient has been instructed and is ready to begin the test, set
modulation in position “1 dB” and press the Tone Switch to start the test
proper. The presentation counter will now count the number of presentations
to the patient and will automatically stop after 20 presentations. At the end of
the test the score may be read off on the display.
Note That The
Modulation Cannot
Be Heard Via
Monitor Headset!
MIDIMATE 622 OPERATION MANUAL
8. Special Tests
49
Содержание MADSEN Midimate 622
Страница 1: ...Midimate 622 Clinical Diagnostic Audiometer User Manual Doc no 7 25 9000 09 Part No 7 25 900 0459 ...
Страница 16: ...10 MIDIMATE 622 OPERATION MANUAL 2 Installing Midimate 622 ...
Страница 18: ...Fig 3 The Selection Tree MIDIMATE 622 OPERATION MANUAL 3 General Description 12 Menu Menu Menu ...
Страница 54: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 7 Speech Audiometry 48 ...
Страница 70: ...This page has been intentionally left blank MIDIMATE 622 OPERATION MANUAL 10 Power On Options 64 ...
Страница 74: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 12 Calibration 68 ...
Страница 101: ...ZODIAC 901 Connection To 74 ZODIAC 901 Printout To 59 73 MIDIMATE 622 OPERATION MANUAL Subject Index 95 ...