4. Press Tone Switch to initiate test and pure tones will be introduced
simultaneously to both ears.
TONE
5. Press
MENU
to exit.
MENU
Please refer to section 8.3 for further details on this test.
5.4.4 RAINVILLE TEST
The Rainville Test is used to determine the bone conduction threshold for
one ear without testing the other ear.
1. Follow steps 1-3 and 5 as described in 5.2.2 on page 24.
2. To invoke this test, access the Select Special Test menu and press
Softkey 4. This setup routes pure tone signal in the stimulus channel to
the right earphone, and selects narrow-band noise to the bone
conductor in the masking channel. The following display appears:
SOFTKEY 4
Freq.
Air
R:Tone Bone
NBN
1000 Hz
C/P/W
20 dB
On/Off
20 dB
3. Place the right earphone on the ear to be tested and the bone conductor
on the mastoid bone behind the ear.
4. Adjust HL to desired level and press Tone Switch to initiate test.
TONE
5. Press
MENU
to exit the Rainville Test and return to the Select Function
Menu.
MENU
Please refer to section 8.4 for further details on the Rainville Test.
MIDIMATE 622 OPERATION MANUAL
5. Operating in Test Mode
36
Содержание MADSEN Midimate 622
Страница 1: ...Midimate 622 Clinical Diagnostic Audiometer User Manual Doc no 7 25 9000 09 Part No 7 25 900 0459 ...
Страница 16: ...10 MIDIMATE 622 OPERATION MANUAL 2 Installing Midimate 622 ...
Страница 18: ...Fig 3 The Selection Tree MIDIMATE 622 OPERATION MANUAL 3 General Description 12 Menu Menu Menu ...
Страница 54: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 7 Speech Audiometry 48 ...
Страница 70: ...This page has been intentionally left blank MIDIMATE 622 OPERATION MANUAL 10 Power On Options 64 ...
Страница 74: ...This page has intentionally been left blank MIDIMATE 622 OPERATION MANUAL 12 Calibration 68 ...
Страница 101: ...ZODIAC 901 Connection To 74 ZODIAC 901 Printout To 59 73 MIDIMATE 622 OPERATION MANUAL Subject Index 95 ...