Moxi GO™ User Guide 

 

Page 1 

Introduction

 

The Moxi GO™ from Orflo Technologies is a benchtop instrument that utilizes a combination of 
a 532 nm laser (MXG001 model) or 488nm laser (MXG002 model), one photomultiplier tube, 
Coulter Principle-based cell measurements, and on-board software to provide easy-to- run 
applications and data analysis. The 532nm system comes with a built-in 561nm/LP PMT 
emission filter, ideal for phycoerythrin (PE) conjugated antibodies, propidium iodide (PI) viability 
staining.  The 488nm system comes with both a 525/45nm PMT emission filter (ideal for FITC, 
GFP and other similar dyes) and a 561nm/LP filter (ideal for PE and PI).  The filters are 
swappable depending on the sample to be run. The instrument relies on disposable cassettes 
for sample handling, which alleviates the need for flow cell cleaning and fluidics maintenance 
and the instrument is small enough to be portable between the lab bench and the hood.  

The Moxi GO™ enables highly accurate cell counts and quantitative assessments of cell 
viability, apoptosis, other labeled antibody markers along with multiplexed bead-based assays 
for protein and cellular analysis.

 

Data acquisition occurs within approximately 10 seconds per 

test.  Test results are displayed in the form of histograms and scatter plots, and quantitatively 
show cell number, cell percentage(s) gated, and mean fluorescent intensity value(s) in a 
graphical display.  The instrument can store up to 16 Mb of data and the data may be 
downloaded to a PC or Mac via USB connectivity.

 

IMPORTANT NOTE

The Moxi GO™ is intended for research use only. It is not intended for use in medical diagnostic 
or therapeutic procedures, and it has not been approved or cleared by the Unites States Food 
and Drug Administration for such uses.   

About the

 

User Guide

 

The 

Moxi GO™ User Guide

 provides detailed information for operating, maintaining, and 

troubleshooting the Moxi GO™, including the system, the Cassettes, and application workflows.  
This includes information on how to analyze test results using the onboard analysis features of 
the software.  

The user guide does not include instructions for analyzing the provided/output flow cytometry 
standard (FCS, version 3.1) files offline in third party software.   

Similarly, the user guide does not include detailed protocols for preparing samples prior to 
running them on the instrument.  The results of the assays are dependent upon the proper use 
of reagents and the instrument.  

 

 

Содержание MOXI GO

Страница 1: ...For Research Use Only Not For Use In Diagnostic or Therapeutic Procedures Moxi GO USER GUIDE...

Страница 2: ...Disclaimer The information in this document is subject to change without notice and should not be construed as a commitment by Orflo Neither Orflo nor any of its affiliated corporations assumes respon...

Страница 3: ...made available to the user by selecting the Settings icon 8 Date and Time 8 Passcode Lock 8 Fluorescence Gain 9 Auto Noise Find 9 Acquiring and Analyzing a Sample 10 General Sample Preparation Guideli...

Страница 4: ...aved Tests 31 Compare Overlay Saved Tests 32 Exporting Data Connecting Via USB USB on the go 34 System Information 35 Instrument Firmware and Software Upgrades 35 Updating the Firmware 36 Moxi GO Apps...

Страница 5: ...n and cellular analysis Data acquisition occurs within approximately 10 seconds per test Test results are displayed in the form of histograms and scatter plots and quantitatively show cell number cell...

Страница 6: ...LOW AND AS INDICATED THROUGHOUT THE USER GUIDE BEFORE OPERATING THE SYSTEM Electrical Safety WARNING To avoid the danger of electric shock including the possibility of permanent injury or death Prior...

Страница 7: ...the Moxi GO to service or adjust the laser The instrument has no user serviceable parts and attempts to adjust or service the laser may result in hazardous radiation exposure All service must be perf...

Страница 8: ...r one pack of 25 MF S Cassettes a Quick Start Reference Guide and Registration Card The following figure shows the main visible components of the Moxi GO and the table provides a description of each c...

Страница 9: ...andling the samples Each cassette contains two fill ports one at each end allowing for two tests to be run per cassette Moxi GOTM Accessories The Moxi GO uses a USB Cable and Power Adapter for power s...

Страница 10: ...ge To begin remove Moxi GO USB cable and power adapter from the inner white box Plug the mini USB end of the USB Cable into the instrument Plug the opposite end of the USB Cable standard USB into the...

Страница 11: ...GO Press the Power Button to turn on the instrument The Home Screen will become available Home Screen NOTE The home page appearance may change as additional applications are launched and new icons ar...

Страница 12: ...o set the lock code select the Settings icon and then use the left and right arrows on the screen to highlight the Passcode Lock field Use up or down arrows to change Passcode Lock field value to On a...

Страница 13: ...ight arrows on the screen to highlight the Fluorescence Gain field Use up or down arrows to change and select the field value and then select Done Note the user can also make a one time per test chang...

Страница 14: ...s ml 2 Cellular Assay Kits Please refer to specific Assay Kit Data sheet iii Fluorescence Detection Beads 1 Please refer to specific Assay Kit Data sheet Optimal concentration ranges up may vary sligh...

Страница 15: ...e a cassette with ORFLO lettering side up into the cassette tray either end 1 or 2 Release the Cassette Tray Lever and close the Door Note Should this result in an error make sure that a your cassette...

Страница 16: ...es While the system waits for sample fluid detection an Analyzing message is indicated in top left of the screen Once the fluid is detected the system message changes to Counting and the results are d...

Страница 17: ...commended volume required for a test is 75 L 2 Select the Open Flow Cytometry application from the Home screen image below left 3 For the Open Flow Cytometry application an option of selecting a fluor...

Страница 18: ...ell cycle assay on Moxi GO has been configured to run specifically with the Orflo Technologies LLC Cell Cycle Protocol thereby insuring that the sample fluorescence intensity is appropriately matched...

Страница 19: ...useful for batch analysis mode in which multiple treated samples will be compared to a single Control file Following selection of the Control file the system will make sure that a new cassette has be...

Страница 20: ...M region is 2x the fluorescence ratio of the G0 G1 region The calculated cell counts percentages MFIs and MFI ratios to the G0 G1 region MFI for each cell cycle region are presented in the table below...

Страница 21: ...Moxi GO User Guide Page 17...

Страница 22: ...main test output formats 1 Fluorescence vs Size dot plot OR 2 Size or fluorescence channel histograms Fully labeled images are provided below for each type of output More detailed workflow instruction...

Страница 23: ...Moxi GO User Guide Page 19 Histograms...

Страница 24: ...n Define the analysis region by adjusting the size blue gates Gate the test results based on fluorescent intensity by adjusting the fluorescence red gate markers Adjust the noise yellow gate to exclud...

Страница 25: ...2 25 m Rescaling the X axis is recommended to achieve finer resolution of smaller particle populations and potentially increased separation between particle sub populations CAUTION Once a test has bee...

Страница 26: ...ot Touching the marker at the base of the left gate toggles the marker from red locked image below right to green unlocked images below left and below middle and back Moving the left cursor with base...

Страница 27: ...cent red gate functionality The user can then touch and drag the red markers to adjust the positioning of the fluorescent gate up or down All data points that are located above the fluorescent gating...

Страница 28: ...s noise As an example the image below right shows the edited saved test for the noise setting defined in the image below middle These points are also not included in the population statistics e g tota...

Страница 29: ...ter Plots Fluorescent and Size histogram overlays on the dot plot can be toggled on and off by pressing the blue histogram button at the bottom right of the scatter plot see image to right With histog...

Страница 30: ...Moxi GO User Guide Page 26 Changing Display Axis for Analysis PMT vs Size Histogram Only Views...

Страница 31: ...the X Y icon is pressed the available channels for each axis are displayed in the gray boxes at the top and to the right of the current scatter image above frame at top middle plot The user can select...

Страница 32: ...population and above Upper Population the red fluorescent gate but within the size gated region The percentage of events that fall above below the fluorescence red gate is displayed in the yellow box...

Страница 33: ...ize Histogram Files within test types are auto enumerated by the system e g Via 001 Via 002 with the lowest available number for that test type Files can be renamed at any point immediately after the...

Страница 34: ...ing Yes will remove any history of the test Selecting No cancels the delete operation All or groups of saved tests can be deleted as a batch by selecting the Delete All button in the Saved Data area o...

Страница 35: ...g vs descending Test files and screenshot files can be opened by selecting the green or dark gray row An example is shown in the image below frames from top middle to bottom left on touching the Open...

Страница 36: ...s Moxi GO software provides the functionality to directly compare two tests by overlaying the data from each test in a dot plot or histogram view To compare or overlay data select Saved Data icon from...

Страница 37: ...elected instead click on the red X in the file checkbox Be careful not to choose Delete Selected or data files will be erased The criteria for comparing two tests is as follows The tests need to be of...

Страница 38: ...corresponding overlay for the other PMT if the channel was recorded select the PMT icon One may also view a size histogram overlay of the two sets by touching the Size Hist icon Summary data for the...

Страница 39: ...t the build date as well as the Bluetooth ID number see image below This information is required by ORFLO Technical Support when assisting users with troubleshooting issues 4 Press the Done icon to ex...

Страница 40: ...into a firmware update mode This can be done in either of the following two ways a With the unit powered off quickly press the power reset button five times reminder the unit needs to be in the off po...

Страница 41: ...cations can be entered by selecting the Apps icon on the Home screen Included applications in version 1 29 are Clock Programmable timer with sound notification Loader Puts unit in Firmware OS loader m...

Страница 42: ...ted and or sticky samples Dilute the sample further and or try to further dissociate the sample with pipette trituration and or protease treatment Low protein expression For poorly expressed proteins...

Страница 43: ...for the fluorescence histogram and top bottom for the size histogram For size histograms not overlayed use the white black box background arrows to adjust the vertical scale to properly scale the popu...

Страница 44: ...Contact error If this affects multiple cassettes please contact ORFLO Technical Support Test Error Fluid was not Detected Sample not loaded or volume too small Make sure sample is loaded and volume i...

Страница 45: ...rried out by authorized personnel only Charging the Battery The Moxi GO contains a 3 7 V lithium ion battery that can be charged for approximately 500 cycles The battery can be charged continuously wi...

Страница 46: ...as follows 10 Bleach Clorox or similar cleaning cloths Mild soap detergent Do not clean the instrument with more aggressive solutions CAUTION When sanitizing do not use solutions containing ammonia su...

Страница 47: ...s platforms Measurement Modes Open Platform Pre Defined Assays Yes Open Flow Application available allowing users to define their own protocols and analysis Requires fluorochromes compatible with abov...

Страница 48: ...is only available for purchase in the Unites States You can purchase ORFLO products on line at www orflo com See the Technical Assistance section for information about contacting ORFLO directly Produ...

Страница 49: ...sary of the date of purchase The extended warranty extends the duration of your warranty coverage from the date of original purchase from ORFLO or an approved sales distributor but does not alter chan...

Страница 50: ...ppearing in ORFLO s product literature may not be altered except by expressed written agreement signed by an officer of Orflo Technologies LLC Representations oral or written which are inconsistent wi...