Updating the Firmware
The firmware can be updated through a USB connection. Generic update instructions are
provided below. Specific instructions for the latest version of software will be included in the
release notes for that version, which can be found along with the Software/Firmware download.
For updates, the user must connect to a computer’s USB port via the provided USB cable.
1. Download and extract the firmware update files (provided as a .zip file). The number
and file names can vary with each firmware update. Typical files included with the
update include: MANIFEST.TXT, moxiprog.hex, runonce.sh, and KERNEL.COF. Note
the location of the files for later use.
2. Turn off the instrument by pressing the power button.
3. Unplug the power adapter to expose the USB connector for the computer. Plug the Moxi
GO™ into the computer that has the downloaded firmware update files.
4. Next, the user needs to put the instrument into a firmware update mode. This can be
done in either of the following two ways:
a. With the unit powered off, quickly press the
power/reset button five times (reminder: the unit
needs to be in the off position to start this
process). When successful, the screen will
display with text under the ORFLO logo that
indicates that the system has entered the
“ORFLO Load” (see image to the right/top). If
unsuccessful, turn the power off and retry.
b. From the
Apps
screen (off of the
Home
screen)
press the
Loader
app icon
5. The Moxi GO™ will show up as an external drive on the
computer (on Mac use the Finder, on PCs use the
Windows Explorer to see it).
6. Drag all extracted files onto the Moxi GO drive. The
Moxi GO™ screen will update to show a “USB File
Transfer” message with an
OK
(see image right/bottom).
7. Once the file copy has completed, safe-eject the disk
from your computer.
8. Touch the
OK
button to proceed. The unit will then load
the new firmware (showing a status bar) and
automatically power-off.
9. The firmware update is now complete. Unplug the unit
from the computer and plug it back into the AC adapter. The unit has been updated and
is now ready for use.