64
4-1 Restrictions
If the DIP switch setting is changed to NON VISUAL, the Controller's settings
can be changed using the DIP switch without having to connect a monitor or a
console.
Notice
Turn OFF the power supply before changing the DIP switch settings.
The restrictions that apply when using the Z300 in Non-visual Mode are pro-
vided below.
- Measurement conditions set using the monitor and console are disabled.
- Even if a monitor and console are connected, console operations will have
no effect.
- Only 1 Sensor can be connected. It must be connected to SENSOR 0.
- Measurement is based on the following conditions.
- Sampling time:
0.96 ms
- Smoothing:
None
- Noise removal:
None
- Output:
Analog output only
- Input:
LD-OFF input only
- The settings that can be changed are as follows.
- The Sensor mounting status can be selected.
- The Sensor's sensitivity can be set.
- The response time for analog outputs can be selected.
- The output method for measurement failure can be selected.
Refer to page 55 for details of setting methods.
Restrictions
Section 4-1
Содержание Z300 V3 - SETUP
Страница 3: ...2 ...
Страница 7: ...6 ...
Страница 21: ...20 ...
Страница 59: ...58 Long range Model Z300 S10 Super Long range Model Z300 S60 Mounting the Sensor Section 3 2 ...
Страница 62: ...61 Z300 S10 Z300 S60 Mounting the Sensor Section 3 2 ...
Страница 63: ...62 ...
Страница 69: ...68 DIP Switch Settings Section 4 2 ...
Страница 72: ...71 SECTION 5 Maintenance This section explains maintenance procedures for ensuring the Z300 s performance ...
Страница 77: ...76 ...
Страница 80: ...79 Z300 S10 Z300 S60 42 8 12 5dia 42 8 12 5dia Sensor Section 7 1 ...