CP TD1 Reference Manual V 1.44
120
Test current (RMS, selective)
Test voltage (RMS, selective)
Capacitance Cp (equivalent parallel circuit)
Dissipation factor DF (tan
δ
)
f0 ± 10Hz
1.2s
> 110dB at fx = f0 ± (10Hz or more)
f0 ± 20Hz
0.9s
> 110dB at fx = f0 ± (20Hz or more)
Terminal
Range
Resolution
Typical accuracy
Conditions
IN A or
IN B
1
1. IN A (red) or IN B (blue), depending on the mode.
0 ... 5A AC
5 digits
error < 0.3% of reading
+ 100nA
Ix < 8mA
error < 0.5% of reading Ix > 8mA
Range
Resolution
Typical accuracy
0 ... 12000V AC
1V
error < 0.3% of r 1V
Range
Resolution Typical accuracy
Conditions
1pF ... 3µF 6 digits
error < 0.05% of
r 0.1pF
Ix < 8mA,
Vtest = 300V ... 10kV
error < 0.2% of reading
Ix > 8mA,
Vtest = 300V ... 10kV
Range
Resolution Typ. accuracy
Conditions
0 ... 10%
(capacitive)
5 digits
error < 0.1% of reading
+ 0.005%
1
1. Reduced accuracy of DF at mains frequency or its harmonics. Mains frequency suppression
available by precisely selecting a mains frequency of *50Hz or *60Hz in the "Hz" column.
f = 45 ... 70Hz,
I < 8mA,
Vtest = 300V ... 10kV
0 ... 100
(0...10000%)
5 digits
error < 0.5% of reading
+ 0.02%
Vtest = 300V ... 10kV
Содержание CP TD1
Страница 1: ...CP TD1 Reference Manual...
Страница 14: ...CP TD1 Reference Manual V 1 44 14...
Страница 20: ...CP TD1 Reference Manual V 1 44 20...
Страница 32: ...CP TD1 Reference Manual V 1 44 32...
Страница 38: ...CP TD1 Reference Manual V 1 44 38...
Страница 84: ...CP TD1 Reference Manual V 1 44 84 Figure 6 14 Frequency scan for H L f 5 kV...
Страница 114: ...CP TD1 Reference Manual V 1 44 114...
Страница 128: ...CP TD1 Reference Manual V 1 44 128...
Страница 154: ...CP TD1 Reference Manual V 1 44 154 Figure 12 12 Diverter switch for 110 kV 40 MVA...
Страница 184: ...CP TD1 Reference Manual V 1 44 184...
Страница 186: ...CP TD1 Reference Manual V 1 44 186...