CP TD1 Reference Manual V 1.44
24
If you experience substantial deviations, re-calibrate the
CP TD1
:
•
Cx = Cref / Cmeas and
•
DF/PF + = DFref - DFmeas
as described in 3.3 on page 29.
A re-calibration of the
CP TD1
is also shown in the test report (.xml file).
Note:
If you change the factory-made calibration, the responsibility for the
accuracy of the
CP TD1
will be in your hands.
2.4.1
Calibration Tips
•
For calibration set the averaging factor to maximum and the filter bandwidth
to ± 5Hz (refer to 3.1 ”TanDelta Test Card - Main Page (1/2)” on page 25).
•
To reset to the factory settings, select "DF/PF+" to 0.0 ppm and "Cx" to 1.000
(refer to 3.3 ”TanDelta Test Card - Settings Page (1/2)” on page 29).
Содержание CP TD1
Страница 1: ...CP TD1 Reference Manual...
Страница 14: ...CP TD1 Reference Manual V 1 44 14...
Страница 20: ...CP TD1 Reference Manual V 1 44 20...
Страница 32: ...CP TD1 Reference Manual V 1 44 32...
Страница 38: ...CP TD1 Reference Manual V 1 44 38...
Страница 84: ...CP TD1 Reference Manual V 1 44 84 Figure 6 14 Frequency scan for H L f 5 kV...
Страница 114: ...CP TD1 Reference Manual V 1 44 114...
Страница 128: ...CP TD1 Reference Manual V 1 44 128...
Страница 154: ...CP TD1 Reference Manual V 1 44 154 Figure 12 12 Diverter switch for 110 kV 40 MVA...
Страница 184: ...CP TD1 Reference Manual V 1 44 184...
Страница 186: ...CP TD1 Reference Manual V 1 44 186...