16
SPECTANO 100 User Manual
6.2
DTS1
dielectric test sample box
Figure 6-3
DTS1
dielectric test sample box
Death or severe injury caused by high voltage or current
►
Use the
DTS1
box only together with the
SPECTANO 100
test device and
the delivered accessories and cables.
►
Read the safety instructions in 2 "Safety instructions" in this User Manual
before using the
DTS1
box together with the
SPECTANO 100
test device.
Equipment damage possible
►
Do not supply a continuous input current > 50 mA
peak
. An overload at the
►
SPECTANO 100
input can cause damages of the device.
►
Always configure the max. test output voltage and the input test
frequency for the connected test object so that the maximum input current
will not be reached.
►
For information on how to configure the output signals for testing with the
►
DTS1
dielectric test sample box, see “Technical data” in the SPECTANO
100 User Manual.
Sample 1
to connect to the
INPUT
connector of
SPECTANO 100
Input
connector (signal source input)
Max. continuous input current: 50 mA
peak
to connect to the
OUTPUT
connector of
SPECTANO 100
Sample 2
RC circuit (for more information, see
the front panel of the
DTS1
box) to
connect to the
INPUT
connector of
SPECTANO 100
DANGER
NOTICE