54
SPECTANO 100 User Manual
Table 10-10
Test cell settings (continued)
Data
Description
Outer diameter of sample
(d
2
)
Outer diameter of the material sample
Guard gap (g)
Gap of the guard ring
Electrode length (l)
Length of the test cell’s electrode
1. Depends on the test cell type selected under
Test cell configuration
.
The following table describes the sweep settings.
Table 10-11
Sweep settings
Data
Description
Test modes
Select a test mode
Select the measurement mode for your test:
•
FDS (only)
: Frequency Domain Spectroscopy
(5 μHz…5
kHz)
•
PDC (only)
: Polarization and Depolarization
Current Method (20 μHz…99
mHz)
•
Combined
: Combines the FDS and PDC
methods in one measurement
(20 μHz…5 kHz). The PDC method is used for
frequencies below the set limit by transforming
the PDC results to FDS results to reduce the
measurement time up to 75% compared to
frequency domain measurements (FDS).
PDC to FDS conversion
1
Use PDC method for
frequencies below
Switching frequency between the FDS and PDC
methods
Convert FDS from PDC
data until
Defines the low-frequency limit of the
transformation from PDC to FDS.
Do not re-calculate stable
PDC data
Select the
Do not re-calculate stable PDC data
check box if you do not want to recalculate the
PDC data when the data does not change with the
increasing time of the PDC measurement.
Sampling points
Points generation
Select
Auto
to generate default frequency sweep
list.
Select
Custom
to specify your own frequency
sweep list.
Start frequency
Start frequency of frequency sweep
Stop frequency
Stop frequency of frequency sweep
FDS points per decade
5
Sample points per decade when using the FDS
method