![Omicron Lab SPECTANO 100 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/omicron-lab/spectano-100/spectano-100_user-manual_3978180058.webp)
57
SPECTANO Analyzer Suite
Table 10-12
Reference settings (continued)
Data
Description
Noise suppression
Select the
Standard
or
Improved
option for the
suppression of the noise interference. If you
select
Improved
, the noise-to-signal ratio and,
consequently, the measurement accuracy will be
improved but the measurement time will increase.
Pre-measurement checks
Check for noise and
connectivity problems
Select the
Check for noise and connectivity problems
check box to check for noise and connectivity
problems before starting a test.
Delay the start of the measurement until
The depolarization current
falls below
Select this option and enter the threshold
depolarization current to delay the start of the
measurement until the depolarization current falls
below the threshold.
The time elapsed
Select this option and enter the time limit to delay
the start of the measurement until the time limit
elapsed.
1. Only available for test cells with disk electrodes
2. For the specified geometry of the electrodes,
SPECTANO Analyzer Suite
calculates the vacuum capacitance automatically. If you
select
Others
in the test cell settings, enter the vacuum capacitance manually.
3. Only available for the PDC and combined measurement modes selected in the sweep settings under
Measurement mode
.
4. Only available for the FDS and combined measurement modes selected in the sweep settings under
Measurement mode
.
The following table describes the measurement settings.
Table 10-13
Measurement settings
Data
Description
General measurement settings
Temperature
Temperature of the test cell/sample
Humidity
Environmental relative humidity
Material sample settings
Material type
Type of the material sample
Serial number
Serial number of the material sample
Sample thickness
Position A
1
Sample thickness at position A
Sample thickness at position B
Sample thickness at position C
Sample thickness at position D