45
SPECTANO Analyzer Suite
Cylindrical
electrodes
Other
geometries
Cylindrical electrodes and other geometries
In this case, perform one reference measurement:
1. In the Navigation pane, click
Reference measurement(s)
, and then click
Graph and result table
. On the ribbon, the
Reference(s)
tab is selected
automatically.
2. Click the
Start reference
button
on the ribbon, and then click
Start
next
to
Reference 1
.
3. The reference measurement is running, and the measurement results are
displayed in the graph view.
4. After the reference measurement has finished, the
Start reference
button
displays the status performed .
5. Now, you can start the correction procedure (see 10.4.2 "Correction
procedure" later in this chapter).
Note:
You can perform the reference measurements before or after the
material measurements.
The
Reference measurement(s)
section and the automated correction
procedure is not available for other geometries. To determine the vacuum
capacitance c
0
read chapter 13.8 "Other geometries” “Vacuum capacitance c
0
and correction procedure”.