51
SPECTANO Analyzer Suite
10.6.3 Result view controls
With the control options on the right of the result view workspace, you can control how the
measurement results are displayed. The following table describes the result view controls.
Table 10-8
Result view controls
Control
Description
Chart result view
Data
Bode: Tanδ
Bode diagram displaying the tangent delta (tanδ)
versus frequency
Bode: R and X
1
Bode diagram displaying the resistance and
reactance versus frequency
Bode: c’ and c’’
Bode diagram displaying the complex
capacitance versus frequency
Bode: ε
r
’ and ε
r’’
Bode diagram displaying the complex relative
permittivity versus frequency
Bode diagram displaying the absolute value and
phase of the impedance versus frequency
Bode diagram displaying the absolute value and
phase of the admittance versus frequency
PDC: Ipol and Idepol
2
Polarization current, depolarization current,
polarization index, specific
3
and electrical
resistance or conductivity in time domain
Cole-Cole diagram displaying complex
impedance
Cole-Cole diagram displaying complex
capacitance
Cole-
Cole: Complex ε
Cole-Cole diagram displaying complex relative
permittivity
Scaling
x-Axis
Select the scaling of the x-axis (linear or
logarithmic).
y-Axis
Select the scaling of the y-(linear or
logarithmic).
Left y-Axis
Ymax
Select the maximum value of the left y-axis.
Ymin
Select the minimum value of the left y-axis.
Right y-Axis
Ymax
Select the maximum value of the right y-
axis.
Ymin
Select the minimum value of the right y-
axis.