12
Technical data
12.1
Output
Voltage range
±100 mV to ±200 V
peak
Max. current
50 mA
peak
12.2
Input
FDS
1
input resistance
<2 Ω
PDC
2
input resistance
2 kΩ
FDS measurement current
≤50 mA
peak
PDC measurement current
≤10 mA
1. Frequency Domain Spectroscopy
2. Polarization Depolarization Current (Time Domain Method)
12.3
Frequency range of
3. This mode combined the FDS and PDC method in one measurement.
12.4
Measurement range
4
Impedance
100 Ω to 20 TΩ
Capacitance
10 pF to 100 µF
Tanδ
≤
3x10
-4
4. Depends on chosen measurement mode, voltage and frequency.
FDS mode
5 µHz to 5 kHz
PDC mode
20 µHz to 99 mHz
Combined
3
mode
20 µHz to 5 kHz