CT SB2 User Manual
28
Tap in Use
Select the tap combination actually used during operation of
the CT (e.g. X1-X5) using the soft keys. By default, the "Tap
in Use" field is set to the full tap combination.
For the tap combination selected here, the
CT Analyzer
displays the detailed test results in the
Resistance
,
Excitation
and
Ratio
cards. The automatic test
assessment in the
Assessment
card is however always
done for the full tap combination given by the
Number of Taps
(e.g. for X1-X5 if number of taps = 5).
Taps
This column lists all possible tap combinations of the CT
(e.g. X1-X5, X1-X4, X1-X3, ...). The number of available tap
combinations depends on the number of taps specified in
the "Number of Taps" field.
Note:
When the
MR-Config.
card’s tab is highlighted you
can select whether X1 or the highest available tap is used as
the common tap. The default common tap can be selected in
the device settings of the
CT Analyzer
, see chapter 5 on
Using the
Show Inter Taps
soft key you can display the
intertap combinations instead of the tap combinations. If the
intertap combinations are displayed, the soft key changes to
Show Taps
.
Ipn : Isn (A)
Enter the nominal ratio for each single tap combination.
If you selected a predefined multi-ratio scheme for
IEEE C57.13 in the
CT-Object
card, this column is
automatically filled with predefined ratios for all tap
combinations.
If you entered I
pn
and I
sn
manually in the
CT-Object
card,
you have to specify the I
pn
for each tap combination in this
column. I
sn
is always taken from the
CT-Object
card.
Note:
The
CT Analyzer
automatically performs a plausibility
check for the ratios entered by the user. For example, an
error message is displayed if the I
pn
entered for X1-X3 is
higher than the I
pn
specified for X1-X4.
Note:
The data for the full tap combination can only be set in
the
CT-Object
card.
Содержание CT Analyzer
Страница 1: ...CT Analyzer CT SB2 User Manual...
Страница 3: ...CT Analyzer User Manual...
Страница 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Страница 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Страница 50: ...CT Analyzer User Manual 48...
Страница 60: ...CT Analyzer User Manual 58...
Страница 80: ...CT Analyzer User Manual 78...
Страница 162: ...CT Analyzer User Manual 160...
Страница 172: ...CT Analyzer User Manual 170...
Страница 188: ...CT Analyzer User Manual 186...
Страница 194: ...CT Analyzer User Manual 192...
Страница 195: ...CT SB2 User Manual...
Страница 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Страница 204: ...CT SB2 User Manual 10...
Страница 206: ...CT SB2 User Manual 12...
Страница 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Страница 212: ...CT SB2 User Manual 18...
Страница 226: ...CT SB2 User Manual 32...
Страница 242: ...CT SB2 User Manual 48...
Страница 248: ...CT SB2 User Manual 54...
Страница 250: ...CT SB2 User Manual 56...
Страница 252: ...CT SB2 User Manual 58...