183
Error and Warning Messages
930.xxx
Error [930.xxx] Measurement input Vsec defective, contact next Omicron service
center.
Reason:
Power supply for analog input "Sec" is defective.
Solution:
Contact your next OMICRON service center.
931.xxx
Error [931.xxx] Measurement input Vprim defective, contact next Omicron service
center.
Reason:
Power supply for analog input "Prim" is defective.
Solution:
Contact your next OMICRON service center.
932.xxx
Error [932.xxx] Data error.
Reason:
The sequence of reading of the analog input channels is not correct.
Solution:
If this error occurs more frequently, contact your next OMICRON service
center.
933.xxx
Error [933.xxx] Hardware failure, contact next Omicron service center.
Reason:
Temperature detection circuit on secondary side is defective.
Solution:
Send the device to your next OMICRON service center.
Safety action: Power supply is switched off and all measurements are disabled.
934.xxx
Warning [934.xxx] Reverse power, don't disconnect any wires, don't switch off mains,
wait until power is dissipated. I = xxxxA.
Reason:
The device receives reverse power of > 20mA.
Solution:
Wait until the power is dissipated within the device and the error message
disappears.
Safety action: All measurements are stopped until reverse power is dissipated.
935.xxx
Error [935.xxx] No valid CT Analyzer software. Insert a CF-Card with valid Software and
press "Update Firmw.".
Reason:
Cannot find valid
CT Analyzer
software.
Solution:
Insert a Compact Flash card with valid software (
CTAnalyzer.bin
) in the
directory
A:\Omicron
.
936.xxx
Warning [936.xxx] Corrupt Calibration Data for voltage inputs. Change to tools menu
and try to reset factory calibration. Until update of calibration data the device will be not
calibrated.
Reason:
The calibration data checksum for the analog inputs is not correct.
Solution:
Try to update the factory calibration using the tools menu. If this does not solve
the problem, contact your next OMICRON service center.
937.xxx
Warning [937.xxx] Corrupt Calibration Data for power output. Change to tools menu
and try to update factory calibration. Until update of calibration data the device will be
not calibrated.
Reason:
The calibration data checksum for the power output is not correct.
Solution:
Try to update the factory calibration using the tools menu. If this does not solve
the problem, contact your next OMICRON service center.
938.xxx
Warning [938.xxx] MIF data block 1 corrupt.
Reason:
The checksum of the factory settings data is faulty.
Solution:
Licensing or device settings data are possibly corrupt. Contact your next
OMICRON service center.
939.xxx
Warning [939.xxx] MIF data block 2 corrupt.
Reason:
The checksum of the factory settings data is faulty.
Solution:
Licensing or device settings data are possibly corrupt. Contact your next
OMICRON service center.
Содержание CT Analyzer
Страница 1: ...CT Analyzer CT SB2 User Manual...
Страница 3: ...CT Analyzer User Manual...
Страница 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Страница 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Страница 50: ...CT Analyzer User Manual 48...
Страница 60: ...CT Analyzer User Manual 58...
Страница 80: ...CT Analyzer User Manual 78...
Страница 162: ...CT Analyzer User Manual 160...
Страница 172: ...CT Analyzer User Manual 170...
Страница 188: ...CT Analyzer User Manual 186...
Страница 194: ...CT Analyzer User Manual 192...
Страница 195: ...CT SB2 User Manual...
Страница 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Страница 204: ...CT SB2 User Manual 10...
Страница 206: ...CT SB2 User Manual 12...
Страница 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Страница 212: ...CT SB2 User Manual 18...
Страница 226: ...CT SB2 User Manual 32...
Страница 242: ...CT SB2 User Manual 48...
Страница 248: ...CT SB2 User Manual 54...
Страница 250: ...CT SB2 User Manual 56...
Страница 252: ...CT SB2 User Manual 58...