CT Analyzer User Manual
176
403.xxx
Warning [403.xxx] Check CT SB2 switch box wiring: CTA OUTPUT, TAPS XN and/or X1
may be disconnected.
Reason:
The CTA OUTPUT sockets of the
CT SB2
are not properly connected to the
CT Analyzer
output.
Solution:
Make sure that the CTA OUTPUT sockets on the
CT SB2
are properly
connected to the
CT Analyzer
and restart your test.
404.xxx
Warning [404.xxx] Check CT SB2 switch box wiring: CTA PRIM and/or PRIM IN terminal
may be disconnected.
Reason:
The CTA PRIM sockets of the
CT SB2
are not properly connected to
CT Analyzer
input PRIM.
Solution:
Make sure that the CTA PRIM sockets on the
CT SB2
are properly connected
to
CT Analyzer
input PRIM and restart your test.
405.xxx
Warning [405.xxx] Check CT SB2 switch box wiring: CTA SEC terminal, TAPS XN IN
and/or X1 IN may be disconnected.
Reason:
The CTA SEC sockets of the
CT SB2
are not properly connected to
CT Analyzer
input SEC.
Solution:
Make sure that the CTA SEC sockets on the
CT SB2
are properly connected
to
CT Analyzer
input SEC and restart your test.
406.xxx
Warning [406.xxx] Check CT SB2 switch box wiring: <text depending on connection
actually detected>
Reason:
The CTA SEC sockets of the
CT SB2
are not properly connected to
CT Analyzer
input SEC.
Solution:
Make sure that the CTA SEC sockets on the
CT SB2
are properly connected
to
CT Analyzer
input SEC and restart your test.
407.xxx
Warning [407.xxx] Check CT SB2 switch box wiring: Wrong polarity at CTA SEC or CTA
OUTPUT terminal
Reason:
Reverse polarity at the CTA SEC sockets or the CTA OUTPUT sockets of the
CT SB2
.
Solution:
Make sure that the CTA SEC sockets and the CTA OUTPUT sockets on the
CT SB2
are properly connected to
CT Analyzer
and restart your test.
408.xxx
Warning [408.xxx] Check CT SB2 switch box wiring: <text depending on missing
connection actually detected>
Reason:
At least one of the CT taps enabled for multi-ratio testing is not connected to
the
CT SB2
correctly.
Solution:
Make sure that all CT taps are connected correctly to the TAPS terminals on
the
CT SB2
and restart your test.
409.xxx
Warning [409.xxx] Check CT SB2 switch box wiring: <text depending on incorrect
connection actually detected>
Reason:
Wiring of at least two CT taps is mixed, e.g. CT tap X2 is connected to TAPS X3
on the
CT SB2
and vice versa.
Solution:
Make sure that all CT taps are connected to the correct TAPS terminals on the
CT SB2
and restart your test.
411.xxx
Warning [411.xxx] Firmware download to switch box failed! Error: SBErrNo
Reason:
CT SB2
firmware download failed.
Solution:
Ensure that the
CT SB2
is connected to the
CT Analyzer
and repeat the
firmware download to the
CT SB2
.
Содержание CT Analyzer
Страница 1: ...CT Analyzer CT SB2 User Manual...
Страница 3: ...CT Analyzer User Manual...
Страница 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Страница 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Страница 50: ...CT Analyzer User Manual 48...
Страница 60: ...CT Analyzer User Manual 58...
Страница 80: ...CT Analyzer User Manual 78...
Страница 162: ...CT Analyzer User Manual 160...
Страница 172: ...CT Analyzer User Manual 170...
Страница 188: ...CT Analyzer User Manual 186...
Страница 194: ...CT Analyzer User Manual 192...
Страница 195: ...CT SB2 User Manual...
Страница 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Страница 204: ...CT SB2 User Manual 10...
Страница 206: ...CT SB2 User Manual 12...
Страница 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Страница 212: ...CT SB2 User Manual 18...
Страница 226: ...CT SB2 User Manual 32...
Страница 242: ...CT SB2 User Manual 48...
Страница 248: ...CT SB2 User Manual 54...
Страница 250: ...CT SB2 User Manual 56...
Страница 252: ...CT SB2 User Manual 58...