179
Error and Warning Messages
813.xxx
Warning [813.xxx] One or more parameters could not be found. Some calculations may
not work.
Reason:
The .xml report file read from the Compact Flash card is of a newer format than
the device can generate.
Solution:
The loaded file is incompatible or the software version on your
CT Analyzer
is
older than the software version that has generated the report. To enable your
CT Analyzer
to read this report, it is necessary to update the
CT Analyzer
software.
814.xxx
Warning [814.xxx] You tried to load an old report. One or more parameters could not
be found. Some calculations may not work.
Reason:
One or more entries could not be found in the .xml report file.
Solution:
The software version on the
CT Analyzer
is newer than the software version
used to generate the report. The report probably does not contain all
parameters supported by the new
CT Analyzer
software.
815.xxx
Error [815.xxx] Folder must be empty. Can't remove folder <Folder Name>.
Reason:
The folder selected for deletion is not empty. Only empty folders can be
deleted.
Solution:
Navigate to the folder you want to delete. Open the folder and delete all
contained files and subfolders. Then you can delete the empty folder.
816.xxx
Error [816.xxx] You tried to load a report built with a <Device>.
Reason:
The .xml file seems to be not a
CT Analyzer
report.
Solution:
The report cannot be loaded by the
CT Analyzer
.
817.xxx
Error [817.xxx] The xml file has no valid OMICRON report style.
Reason:
The .xml file has no valid OMICRON report style.
Solution:
The report cannot be loaded by the
CT Analyzer
.
818.xxx
Warning [818.xxx] Moving of folders not supported.
Reason:
Moving of folders is not supported by the
CT Analyzer
.
Solution:
Select only files to move, not folders. Using the shift key and the up/down
cursor keys it is possible to select any number of files within a folder.
819.xxx
Warning [819.xxx] File name contains an invalid character.
Reason:
An invalid character was used to specify the file name.
Invalid characters: \ / : * ? \ < > |
Solution:
Only use valid characters for the file name.
820.xxx
Error [820.xxx] Memory management error. Contact the next OMICRON service center.
Reason:
Memory management error.
Solution:
Switch the
CT Analyzer
off, wait a second then switch it on again. If this error
occurs more frequently, you should contact your next OMICRON service
center.
821.xxx
Error [821.xxx] Could not update firmware.
Reason:
An error has occurred in the boot loader software.
Solution:
Insert a Compact Flash card with valid software (
CTAnalyzer.bin
) in the
directory
A:\Omicron\
and switch the
CT Analyzer
off and on again.
823.xxx
Error [823.xxx] Could not update user text!
Reason:
An error has occurred in the user text loader software.
Solution:
Insert a Compact Flash card with valid software (
CTUser_xxx.bin
) in the
directory
A:\Omicron\
and try again.
Содержание CT Analyzer
Страница 1: ...CT Analyzer CT SB2 User Manual...
Страница 3: ...CT Analyzer User Manual...
Страница 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Страница 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Страница 50: ...CT Analyzer User Manual 48...
Страница 60: ...CT Analyzer User Manual 58...
Страница 80: ...CT Analyzer User Manual 78...
Страница 162: ...CT Analyzer User Manual 160...
Страница 172: ...CT Analyzer User Manual 170...
Страница 188: ...CT Analyzer User Manual 186...
Страница 194: ...CT Analyzer User Manual 192...
Страница 195: ...CT SB2 User Manual...
Страница 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Страница 204: ...CT SB2 User Manual 10...
Страница 206: ...CT SB2 User Manual 12...
Страница 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Страница 212: ...CT SB2 User Manual 18...
Страница 226: ...CT SB2 User Manual 32...
Страница 242: ...CT SB2 User Manual 48...
Страница 248: ...CT SB2 User Manual 54...
Страница 250: ...CT SB2 User Manual 56...
Страница 252: ...CT SB2 User Manual 58...