User Maintenance
8
User Maintenance
8.1
Care and Cleaning
The
CT SB2
does not require any special maintenance or care. Clean the device
from time to time or as necessary using a cloth dampened with water or
isopropanol alcohol. Always disconnect the
CT SB2
prior to cleaning!
8.2
Replacing Fuses
1. Turn off the
CT SB2
and unplug the power cord.
2. Ground the test object, and disconnect it from the
CT SB2
. By disconnecting
it you prevent a possibly faulty test object from feeding power back into the
CT SB2
.
3. Disconnect the
CT SB2
from the
CT Analyzer
and, if applicable, from the
computer.
4. Locate the blown fuse on the side panel of the
CT SB2
and replace it by an
identical fuse type: T2.0A H 250V (2.0 Amps slow-acting high breaking
capacity wire fuse 5 x 20mm). The
CT SB2
has two fuses of the same type.
Содержание CT Analyzer
Страница 1: ...CT Analyzer CT SB2 User Manual...
Страница 3: ...CT Analyzer User Manual...
Страница 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Страница 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Страница 50: ...CT Analyzer User Manual 48...
Страница 60: ...CT Analyzer User Manual 58...
Страница 80: ...CT Analyzer User Manual 78...
Страница 162: ...CT Analyzer User Manual 160...
Страница 172: ...CT Analyzer User Manual 170...
Страница 188: ...CT Analyzer User Manual 186...
Страница 194: ...CT Analyzer User Manual 192...
Страница 195: ...CT SB2 User Manual...
Страница 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Страница 204: ...CT SB2 User Manual 10...
Страница 206: ...CT SB2 User Manual 12...
Страница 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Страница 212: ...CT SB2 User Manual 18...
Страница 226: ...CT SB2 User Manual 32...
Страница 242: ...CT SB2 User Manual 48...
Страница 248: ...CT SB2 User Manual 54...
Страница 250: ...CT SB2 User Manual 56...
Страница 252: ...CT SB2 User Manual 58...