Specific parameters and settings displayed for IEC 60044-1
metering CTs
The following parameters are only displayed in the
CT-Object
card if the
standard IEC 60044-1 is selected with the type "Metering CT".
Param. Description
Available for
IEC 60044-1
metering CTs,
class
0.1, 0.2, 0.2s, 0.5,
0.5s, 1, 3, 5
FS
Instrument security factor acc. to IEC 60044-1.
Possible values: Integer value from 1 to 30 or soft keys
?
1
,
FS1
,
FS1.5
,
FS2
,
FS5
,
FS10
,
FS20
or
FS30
.
Default: ?.
1. If the question mark is entered for this parameter and a new test is started, the
CT Analyzer
automatically tries to determine the value using its integrated guesser function (see chapter 7 on
page 131).
x
ext. Ipn Extended current rating.
Possible values: 100 to 400% or soft keys
120%
,
150%
,
200%
,
300%
,
400%
.
Default: 120%.
x
Содержание CT Analyzer
Страница 1: ...CT Analyzer CT SB2 User Manual...
Страница 3: ...CT Analyzer User Manual...
Страница 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Страница 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Страница 50: ...CT Analyzer User Manual 48...
Страница 60: ...CT Analyzer User Manual 58...
Страница 80: ...CT Analyzer User Manual 78...
Страница 162: ...CT Analyzer User Manual 160...
Страница 172: ...CT Analyzer User Manual 170...
Страница 188: ...CT Analyzer User Manual 186...
Страница 194: ...CT Analyzer User Manual 192...
Страница 195: ...CT SB2 User Manual...
Страница 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Страница 204: ...CT SB2 User Manual 10...
Страница 206: ...CT SB2 User Manual 12...
Страница 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Страница 212: ...CT SB2 User Manual 18...
Страница 226: ...CT SB2 User Manual 32...
Страница 242: ...CT SB2 User Manual 48...
Страница 248: ...CT SB2 User Manual 54...
Страница 250: ...CT SB2 User Manual 56...
Страница 252: ...CT SB2 User Manual 58...