80
6.3 Insertion of the lamp
CLV-S190 INSTRUCTION MANUAL
Ch.6
8
Turn the knob (B) clockwise by 90
from the horizontal position until it stops while
pushing the knob.
Figure 6.13
9
Turn the knob (A) clockwise by 90
from the horizontal position until it stops while
pushing the knob. (See Figure 6.13)
10
Confirm that the heat sinks are attached firmly by pulling the projection. (See
Figure 6.12)
CAUTION
If the heat sinks are not installed properly, their overheating may result in
equipment damage or drop in the examination light brightness.
11
Store the hexagon wrench back on the rear of
the lamp cover.
Figure 6.14
Lamp cover
Hexagon wrench