88
8.2 Troubleshooting guide
CLV-S190 INSTRUCTION MANUAL
Ch.8
Irregularity description
Possible cause
Solution
The endoscope cannot be
connected to the light
source.
The endoscope is not compatible with
the light source.
Connect an endoscope that is listed in
the “
System chart” on page 93.
The power fails to turn ON.
The power switch is not turned ON.
Turn the power switch ON.
The power cord is not connected.
Connect it to a wall mains outlet as
described in Section 3.6, “Connection
to the AC mains power supply”.
The lamp cover is not closed firmly or
removed.
Close the lamp cover firmly described in
Section 6.3, “Insertion of the lamp”.
The power switch of the mobile
workstation is OFF.
Turn the power switch of the mobile
workstation ON.
The examination lamp does
not ignite.
The examination lamp has not been
ignited yet.
Press the lamp button.
The examination lamp is not installed.
Install an examination lamp as
described in Section 6.1, “Replacement
of the examination (xenon) lamp”.
The examination lamp is not installed
correctly.
Reinstall the examination lamp as
described in Section 6.1, “Replacement
of the examination (xenon) lamp”.
The examination lamp is broken.
Replace the examination lamp with a
new one as described in Section 6.1,
“Replacement of the examination
(xenon) lamp”.
No light is emitted from the
endoscope.
The examination lamp has not been
ignited yet.
Press the lamp button.
The endoscope or light guide is not
connected to the output socket.
Connect the endoscope to the output
socket securely as described in
Section 4.3, “Connection of an
endoscope”.
The examination lamp does
not turn OFF even if the lamp
button is pressed.
The lamp button is pressed only for a
short time.
Press and hold the lamp button for
1 second or more.
The examination lamp does
not ignite, and the
emergency lamp indicator
lights up.
The examination lamp is not installed.
Install an examination lamp as
described in Section 6.1, “Replacement
of the examination (xenon) lamp”.
The examination lamp is not installed
correctly.
Reinstall the examination lamp as
described in Section 6.1, “Replacement
of the examination (xenon) lamp”.
The examination lamp is broken.
Replace the examination lamp with a
new one as described in Section 6.1,
“Replacement of the examination
(xenon) lamp”.
The emergency lamp
indicator blinks.
The emergency lamp has blown or
failed.
Immediately turn the light source OFF,
unplug the power cord from the wall
mains outlet and the power inlet, then
contact Olympus.
Содержание VISERA ELITE CLV-S190
Страница 2: ......
Страница 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Страница 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Страница 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Страница 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Страница 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Страница 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Страница 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...