App.
EMC information
101
CLV-S190 INSTRUCTION MANUAL
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% U
T
(100% dip in U
T
)
for 0.5 cycle/1 cycle
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
Same as left
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of this instrument requires
continued operation during
power mains interruptions, it is
recommended that this
instrument be powered from an
uninterruptible power supply or a
battery.
–
40% U
T
(60% dip in U
T
)
for 5 cycle
70% U
T
(30% dip in U
T
)
for 25 cycle (50 Hz)/
30 cycle (60 Hz)
Phase angle causing
voltage dips: 0
70% U
T
(30% dip in U
T
)
for 25 cycle
0% U
T
(100% dip in U
T
)
for 250 cycle (50 Hz)/
300 cycle (60 Hz)
< 5% U
T
(> 95% dip in U
T
)
for 5 seconds
U
T
is the a.c. mains voltage prior to application of the test level.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
(50 Hz or 60 Hz)
3 A/m
(50 Hz, 60 Hz)
Same as left
It is recommended to use this
instrument by maintaining
enough distance from any
equipment that operates with
high current.
Immunity test
IEC 60601-1-2
(2014)
test level
IEC 60601-1-2
(2007, 2001)
test level
Compliance
level
IEC 60601-1-2 (2007, 2001)
Electromagnetic
environment — Guidance
Содержание VISERA ELITE CLV-S190
Страница 2: ......
Страница 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Страница 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Страница 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Страница 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Страница 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Страница 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Страница 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...