44
4.6 Inspection of the examination light
CLV-S190 INSTRUCTION MANUAL
Ch.4
WARNING
Do not look directly into the distal end of the endoscope, the distal end of the light
guide cable, or the output socket of the light source when they are emitting light.
The intense light may cause eye injury.
4.6
Inspection of the examination light
1
Press the lamp button if the examination lamp
is in standby and confirm that the lamp
indicator
“ON” lights up.
Figure 4.5
2
Confirm that the examination light is emitted
from the distal end of the endoscope. When the
lamp light intensity decrease even if the “500 h”
indicator does not light up, replace the
examination lamp with a new one as described
in Section 6.1, “Replacement of the
examination (xenon) lamp”.
Figure 4.6
3
Press and hold the lamp button for about 1 second: the lamp indicator “STBY” lights
up. (See Figure 4.5)
4
Confirm that the examination light is not emitted from the distal end of the endoscope.
Lamp indicator “ON”
Lamp button
Emergency lamp indicator
Distal end of
endoscope
Examination
light
Содержание VISERA ELITE CLV-S190
Страница 2: ......
Страница 6: ...iv Contents CLV S190 INSTRUCTION MANUAL...
Страница 20: ...14 Summary of the Functions of the Instrument CLV S190 INSTRUCTION MANUAL...
Страница 22: ...16 1 1 Checking the package contents list CLV S190 INSTRUCTION MANUAL Ch 1...
Страница 30: ...24 2 3 Rear and side panels CLV S190 INSTRUCTION MANUAL Ch 2...
Страница 42: ...36 3 6 Connection to the AC mains power supply CLV S190 INSTRUCTION MANUAL Ch 3...
Страница 62: ...56 4 11 After inspection CLV S190 INSTRUCTION MANUAL Ch 4...
Страница 98: ...92 8 3 Returning the light source for repair CLV S190 INSTRUCTION MANUAL Ch 8...