6.3 Insertion of the lamp
79
CLV-S190 INSTRUCTION MANUAL
Ch.6
6
Place heat sink (A) and heat sink (B) so that
their undersides are flat and close the heat sink
clamp firmly.
Figure 6.11
CAUTION
Be sure to tighten the heat sink clamp firmly. Poor heat radiation may result in
equipment damage, examination lamp ignition failure, and a considerable drop in
the life of the examination lamp.
7
Insert both heat sinks (A) and (B) simultaneously into the lamp chamber along the
insertion grooves.
Figure 6.12
Clamp
Heat sink (B)
Heat sink (A)
Heat sink (B)
Knob (A)
Heat sink (A)
Knob (B)
Lamp chamber
Projection
Insertion groove