90
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Inspection of the single use channel cleaning brush
(BW-201B)
1.
Remove the packaging immediately before use (see Figure 7.10).
2.
Confirm that the tip and bristles at the distal end are securely in place.
Check for loose or missing bristles.
3.
Check for bends, scratches and other damage to the shaft.
4.
Check for debris on the shaft and/or in the bristles of the brush.
Figure 7.10
Inspection of the single use channel-opening cleaning
brush (MAJ-1339)
1.
Open the packaging immediately before use (see Figure 7.11).
2.
Check for loose or missing bristles.
3.
Check for bends, scratches, and other damage to the shaft.
4.
Check for debris on the shaft and/or in the bristles of the brush.
Figure 7.11
Tip
Shaft
Bristles
Packaging
Bristles
Handle
Packaging
Shaft