Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
125
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
21.
Using a sterile, lint-free cloth, thoroughly wipe and dry all external surfaces.
High-level disinfection
•
Completely remove all air bubbles from all equipment,
including inner channels. Air bubbles remaining on the
equipment prevent proper disinfection.
•
All disinfection steps should be performed with all equipment
completely immersed. If any part of the equipment is not
immersed completely, disinfectant solution may not
adequately contact all surfaces.
1.
Fill a basin with disinfectant solution at the temperature and concentration
recommended by the disinfectant manufacturer. Use a basin that is deep
enough to completely immerse all equipment.
2.
Immerse all equipment in the disinfectant solution.
3.
Using a lint-free cloth and/or 30 cm
3
(30 ml) syringe, wipe and/or flush all
surfaces with the disinfectant solution to remove all air bubbles.
4.
Using the 30 cm
3
(30 ml) syringe, flush the interior and recessed parts of
them while they are immersed. Ensure that all air bubbles are expelled.
5.
Using the 30 cm
3
(30 ml) syringe, flush the recessed parts of the biopsy
valve with disinfectant solution while the valve is immersed. Ensure that all
air bubbles are expelled (only for reusable biopsy valve (MD-495)).
6.
Rub the bristles of the channel cleaning brushes and channel-opening
cleaning brush to remove all air bubbles.
7.
Using the 30 cm
3
(30 ml) syringe, flush the inside of the suction cleaning
adapter thoroughly with disinfectant solution to remove all air bubbles while
it is immersed in the disinfectant solution.
8.
Cover the basin with a tight fitting lid to minimize the release of disinfectant
vapor. Soak all equipment at the temperature and concentration
recommended by the disinfectant manufacturer.
9.
Remove the suction cleaning adapter from the disinfectant solution. Hold
the end of the adapter and tilt it for expelling the disinfectant solution
remaining in its adapter.
10.
Remove all equipment from the disinfectant solution.