44
Chapter 3 Preparation and Inspection
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Connection of the ultrasonic cable (MAJ-1597) and diagnostic
ultrasound system
Insert the ultrasonic connector properly into the transducer port of the diagnostic
ultrasound system. Rotate the connector’s lock handle 1/4 turn clockwise (see
Connect the Ultrasonic connector to one of the probe
connectors on the lower side (PROBE 3 or 4) of the
ProSound F75 (see Figure 3.24).
If the Ultrasonic connector is connected to one of the probe
connectors on the upper side (PROBE 1 or 2), the operation
panel of the ProSound F75 may hit the Ultrasonic connector,
which may result in the equipment damage. When adjusting
the height, a horizontal and / or vertical position of the
operation panel of the ProSound F75 with the Ultrasonic
connector connected to the PROBE 1 or 2, move slowly the
operation panel with visually confirming the position of the
bottom of the operation panel.