158
Appendix C: Acoustic Output Information in Accordance with the FDA Guidance: “Information for Manufacturers
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Acoustic output table for default settings when combined
with SSD-
5
Transducer Model:
BF-UC180F
Acoustic Output
MI
TIS
TIB
TIC
Maximum Value
0.65
0.31
0.37
0.67
Operating mode
B
B/Bflow
PWD
B/Bflow
Operator
Controls
DVA%
(%)
70
54
70
54
Function
–
–
–
–
B/W Frequency Select
(MHz)
12 MHz
12 MHz
12 MHz
12 MHz
B/W Focus
F4+F7
F6
F6
F6
B/W-Range
(cm)
12.0
12.0
12.0
12.0
B/W Line Density
HIGH
HIGH
HIGH
HIGH
B/W Scan Area
(%)
100
100
100
100
Doppler Image Select
n/a
n/a
PENET
n/a
Doppler Frequency Select
(MHz)
n/a
n/a
5 MHz
n/a
Doppler Focus
n/a
n/a
F7
n/a
Doppler Velocity Range
(cm/sec)
n/a
n/a
24.9
n/a
Color Image Select
n/a
STD
n/a
STD
Color Doppler Frequency Select
(MHz)
n/a
5 MHz
n/a
5 MHZ
Flow Focus
n/a
F7
n/a
F7
Color Velocity Range
(m/sec)
n/a
0.17
n/a
0.17
Color Line Density
n/a
LOW
n/a
LOW
Color Average
n/a
HIGH
n/a
HIGH
Flow Filter
n/a
#5
n/a
#5
Flow Scan Area
(%)
n/a
50
n/a
50
Triplex Velocity Range
n/a
n/a
n/a
n/a
I
SPTA.3
(mW/cm
2
)
8.7
19
67
19
Содержание BF UC180F
Страница 3: ......
Страница 7: ...Contents iv ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF UC180F...
Страница 76: ...Chapter 4 Operation 69 ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF UC180F Figure 4 5 Figure 4 6...
Страница 181: ...174 Appendix D Acoustic Output Information Accordance with IEC 60601 2 37 ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF UC180F...