Chapter 6 Compatible Reprocessing Methods and Chemical Agents
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Steam sterilization (autoclaving)
Ethylene oxide gas sterilization
(use OER-A)
2 – 3.5% glutaraldehyde
70% ethyl or isopropyl
alcohol
Detergent solution
Ultrasonic
cleaning
Water-resistant cap
(MH-553)
Cap chain (MAJ-1723)
2
Channel cleaning
brush (BW-15B)
Channel cleaning
brush (BW-7B)
Channel-opening
cleaning brush
(MH-507)
Suction valve
(MAJ-207)
Biopsy valve (MD-495)
Suction cleaning
adapter (MAJ-222)
Balloon applicator
(MAJ-1352)
Mouthpiece (MA-651)
Ultrasonic cable
(MAJ-1722, MAJ-1597)
compatible
not compatible
Table 6.1
1
The endoscope is compatible with ultrasonic cleaning only when using an
endoscope reprocessor such as OER, OER-A and OER-AW (OER,
OER-A and OER-AW are not available in some areas).
2
The water-resistant caps and the cap chains can only be ultrasonically
cleaned if connected to the endoscope that is being cleaned in an
endoscope reprocessor with an ultrasonic cleaning phase.
3
ACECIDE disinfectant solution is exclusively for OER-A and OER-AW.