Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
83
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
7.1
Required reprocessing equipment
Preparation of the equipment
Prior to cleaning and disinfection or sterilization, prepare the equipment shown in
Use basins that are at least 40 cm by 40 cm (16” by 16”) in
size and deep enough to allow the endoscope to be
completely immersed.
Figure 7.1
• Detergent solution
• Disinfectant solution
• Personal protective equipment
• 500 cm
3
(500 ml) containers
• Large non-metallic basin with a
tight fitting lid for detergent and
disinfectant solution
• Clean, lint-free cloth
• Clean water (see Section 6.2)
• Large basin for rinsing and
leakage testing
• Sterile, lint-free cloth
• Sterile water
• 30 cm
3
(30 ml) syringe
• Sterile, cotton swabs
• 70% ethyl or isopropyl alcohol
• Soft brush
• Small containers
Channel cleaning
brush (BW-7B
1
) or
single use single-ended
cleaning brush
(BW-400B
1
)
Suction cleaning adapter
(MAJ-222)
Channel cleaning brush
(BW-15B) or
single-use channel cleaning
brush (BW-201B, 3 pcs
1
)
Channel-opening cleaning brush
(MH-507) or
single use channel-opening cleaning
brush (MAJ-1339, 3 pcs
1
)
Water-resistant cap
(MH-553, 2 pcs)
Leakage tester
(MB-155)
Maintenance unit
(MU-1)
Suction pump
(KV-4, KV-5, SSU-2)
Single use combination
cleaning brush
(BW-411B
1
)
1
These products may not be available in some areas.