122
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Equipment needed
Prepare the following equipment and wear appropriate personal protective
equipment.
Reusable parts that are normally reprocessed with the
endoscope
Reusable parts that are normally reprocessed separately
from the endoscope
Preparation
Wear appropriate personal protective equipment.
• Personal protective equipment
• Soft brush
• Clean, lint-free cloths
• Sterile, lint-free cloths
• Large basins
• Detergent solution
• Disinfectant solution
• Clean water (nonsterile water)
• Sterile water (for sterile water rinse)
• Small container
• 70% ethyl or isopropyl alcohol
• 30 cm
3
(30 ml) syringe
• Channel cleaning brush or
single-use channel cleaning brush
(BW-15B)
(BW-201B)
• Single use combination cleaning brush
(BW-411B)
• Single use single-ended cleaning brush
(BW-400B)
• Suction cleaning adapter
(MAJ-222)
• Water resistant caps
(MH-553)
• Mouthpiece
(MA-651)
• Channel cleaning brush
(BW-7B)
• Channel cleaning brush
(BW-15B)
• Channel-opening cleaning brush
(MH-507)
• Suction valve
(MAJ-207)
• Biopsy valve (optional item)
(MD-495)
• Balloon applicator
(MAJ-1352)