Chapter 10 Troubleshooting
143
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Ultrasound (display monitor image)
Balloon water feeding and aspiration
Others
Irregularity
description
Possible cause
Solution
Echo appears where
there is no object.
Multiple echoes
generated from an
object with strong
reflection.
Move the endoscope’s distal end
to another position and check to
see if the echo in question moves
to a greater degree than other
echoes. If this happens, multiple
echoes have coalesced and
disturbed the ultrasonic image.
Keep the endoscope’s distal end
away from an object with strong
reflection.
Irregularity
description
Possible cause
Solution
Water feeding to the
balloon and/or
aspiration from the
balloon does not work.
Balloon water supply
channel is dirty.
Reprocess the channel.
Irregularity
description
Possible cause
Solution
The remote switch does
not work.
The wrong remote
switch is operated.
Operate the correct remote
switch.
The remote switch
function has been set
improperly.
Set the remote switch function
correctly as described in the video
system center’s instruction
manual.