13
•
Carefully place the cap and turn it clockwise. Make sure that you do not cross thread it, and tighten
firmly by hand. Do not over-tighten (see image 4 in Figure 12).
•
Connect the Oxygen Analyzer tubes are connected to the Oxygen Analyzer
–
Gas input connector and
Oxygen Analyzer
–
Gas output connector (see image 5 in Figure 12). Make sure that the Oxygen
analyzer tubes are securely connected.
Tip
►
Make sure to push the tubes all the way into the connectors thus avoiding any gas leak.
1
2
3
4
5
6
O2 SENSOR
ORING
Figure 12. O2 Sensor installation.
5.
Figure 13 shows the connections between the N2, Air and CO2 tanks and the respective connectors on the rear
panel of CO2-O2-UNIT-BL [0-10; 0-1].