10
8
Installation
The following paragraph illustrates how to install and to use CO2-O2-UNIT-BL [0-10; 0-1].
Setting up CO2-O2-UNIT-BL [0-10; 0-1]
The following instructions shall give you the possibility to install the unit quickly. For safe operation
of the unit it is absolutely necessary to read carefully all the instructions and safety notes.
1.
Install the supplied pressure gauge between the CO2 supply and CO2-O2-UNIT-BL [0-10; 0-1]. Install the
supplied pressure gauge between the N2 supply and CO2-O2-UNIT-BL [0-10; 0-1]. Follow the arrow on the
rear of the pressure gauge for the correct Gas In-Out (see Figure 7).
GAS IN
GAS OUT
Figure 7. Input and Output gas ports. See symbol on the rear of pressure gauge.
On the bottom of the pressure gauge there is an automatic purging valve. Close it before operating the
pressure gauge. To close the purging valve, looking at the regulator vertically from the top knob, pull it down and
rotate it counterclockwise (see A and B in Figure 8). Now you can operate the pressure gauge.
Purging valve
A
B
Figure 8. How to close the purging valve of the pressure gauge.