6
5
GAS supplied requirements
−
CO2.
CO2 must be available at
1.4 barg (20.3 psig)
, with a connector for 6mm OD rigid tube. Gas source
must be Standard Purity CO2 (coded as 4.5 that means 99.995 % of CO2) and humidity-free gas.
−
N2
. N2 must be available within the pressure range at
1.4 barg (20.3 psig)
, with a connector for 6mm OD
rigid tube. Gas source must be Standard Purity N2 (coded as 4.5 that means 99.995 % of N2) and humidity-
free gas.
−
Air
. Air must be available at
1.4 barg (20.3 psig)
, with a connector for 6mm OD rigid tube.
Note
►
An alternative to the compressed air, OKO-AIR-PUMP-BL allows to use background Air.
6
Compatibility
CO2-O2-UNIT-BL [0-10; 0-1] is compatible with:
•
OKO-TOUCH
(required)
, including:
−
Touch Screen Interface
−
#1 RS232 Serial cable
. To connect the Touch Screen to the unit.
−
MINI-USB-OTG cable
. To connect a USB pen to the OKO-TOUCH for data download.
•
OKO-AIR-PUMP-BL
(
only in case it is required to work in “CO2 in Air” mode)
•
H101-BASIC-BL
•
H101-CRYO-BL
•
H201-T-UNIT-BL
•
H301-T-UNIT-BL-PLUS
•
HM-ACTIVE
•
H101-HM
•
HM-VF
OKO-AIR-PUMP-BL
HM-ACTIVE
H301-T-UNIT-BL
OKO-TOUCH
Figure 2. CO2-O2-UNIT BL [0-10; 0-1] within a typical H301 configuration.