9.2.2.5
Alarms
CO2-O2-UNIT-BL [0-10; 0-1] also allows activating visual and audible alarms.
To set the alarms specifications, press the
Alarms
icon
(see Figure 64 a), then follow the indications
below:
1.
Click on the Gas icon
to set the gas deviation for a period of time then the system triggers on Alarms, as
shown in the Figure 64 b.
Tip
►
Flag the Buzzer checkbox if you want the Alarm to be acoustical as well rather than just being
displayed.
a
b
Figure 64. How to enter in Alarms page (a -b).
2.
Insert the Gas Deviation by scrolling the bar (1 in Figure 65) or pressing the
+/-
icons.
Note
►
The Gas Deviation value defines the tolerance from the Setpoint.
Note
►
The gas deviation value range is 0.1-2.0 %.
3.
Insert the
Gas Time
by scrolling the bar (2 in Figure 65) or pressing the
+/-
icons.
Note
►
The Gas Time value defines the time for which the CO2 / O2 composition may remain outside the
allowed tolerance before CO2-O2-UNIT-BL [0-10; 0-1] triggers in alarm.
Note
►
The alarm time range is 5-20 minutes.
4.
Press
Save
(3 in Figure 65) to confirm.