Figure 34. Offset Calibration - Correction.
Note
►
By pressing Adj CO2 / Adj O2, the system will perform a calibration using the External Probe as a
reference. The offset resulting from the calibration will be displayed in the Offset tab (see Figure 35).
Figure 35. Offset Calibration End.
Note
►
By pressing on Reset the system will delete all custom calibration and restore the factory values (see
Figure 36. Restore the factory values.
9.2.1.2.2
Sensors (with calibration gas)
This function allows performing a calibration using a Span Gas or pure N2 Tank. Before starting the
instrument calibration, make sure you have the Gas Tank (Span Gas or N2 tank)
connected to the port labeled “
N2
”
on the rear panel of CO2-O2-UNIT-BL [0-10; 0-1], then press the
Sensors
icon
, as shown in
Figure 37 a. At this
point several options are available (see Figure 37 b).