![Ohmeda Giraffe OmniBed Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/ohmeda/giraffe-omnibed/giraffe-omnibed_operators-manual_3546691009.webp)
1-5
6600-0342-000 08/08/01
1-5
General Information
Operating Mode Transition
If the canopy is raised while the OmniBed is being operated as an incubator, it
automatically switches to operation as a radiant warmer. If it was in the baby
mode as an incubator, after the canopy is raised it will continue to operate in
the baby mode, using the same baby set temperature you entered earlier.
If the canopy is raised while the OmniBed is being operated in the air mode as
an incubator, it will operate in the manual mode as a warmer. The radiant
heater will activate at the preheat level (the factory set preheat level is 25%;
this can be adjusted on the service screen), and the OmniBed will prompt you
to select a heater power percentage.
If the canopy is lowered while the OmniBed is being operated as a radiant
warmer, it automatically switches to operation as an incubator. If it was in the
baby mode as a warmer, after the canopy is lowered it will continue to operate in
the baby mode, using the same baby set temperature you entered earlier.
If the canopy is lowered while the OmniBed is being operated in the manual
mode as an warmer, it will operate in the air mode as an incubator. The
compartment heater and fan will start at the last air set temperature saved
since power-off, or at the preheat temperature of 33°C, and the OmniBed will
prompt you to select an air set temperature.
Accessory Options
The OmniBed may be equipped with a number of accessory options.
Menu icons of these options are shown on the graphics display at the
right of the control area. Pressing the knob activates the screen. Rotating
the control knob either direction scrolls through each menu icon; highlight-
ing each icon as it is selected. Pressing in the control knob selects the
highlighted option and brings up its menu screen.
?
Options
Done/Exit
Probe 2
36.8
Alarm messages displayed here
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
1. Help
2. Timer
3. Trending
4. Comfort Zone
5. Setup/Customize
6. Scale
7. Day
8. Time*
9. Date*
10. Baby Information
11. Humidity
12. Options
11
2
3
4
5
6
7
1
8
12
CI.31.013
9
10
* The date and time may be set using the
service screen; see the service manual.
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...