![Ohmeda Giraffe OmniBed Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/ohmeda/giraffe-omnibed/giraffe-omnibed_operators-manual_3546691013.webp)
1-9
6600-0342-000 08/08/01
1-9
General Information
Setup/Customize:
Select Setup/Customize icon (wrench) to
bring up the setup screen.
Select “Temp Units” to change temperature displays from the
default Celsius setting to Fahrenheit.
Select “Alarm Volume” to choose from 4 volume levels.
Select “Man Temp Alarm” to choose from 3 settings: choose “Off”
to deactivate the alarm, choose “.5” to set the alarm to activate
when a 0.5°C difference is read between a set temperature and
the baby probe temperature, or choose “1.0” to set the alarm to
activate when a 1.0°C difference is read between a set tempera-
ture and the baby probe temperature. “Off” is the default setting.
The “Man Temp Alarm” can be used in either the warmer Manual
mode or the incubator Air mode. However, to select a set tem-
perature you must first select the Baby mode, input the set point,
then select the Manual or Air modes.
This optional alarm provides you audiovisual notification in the Air
control mode if the skin probe temperature drifts outside the
range you select. For example, if you select 0.5°C on the setup
screen, and then set 36.5°C as the skin temperature threshold in
the Baby mode before switching to the Air mode, the alarm will
sound if the baby’s skin temperature falls below 36°C or rises
above 37°C while in the Air mode. If this alarm is not required, or
if the skin temperature probe is not being used, “Man Temp
Alarm” should always be set to “Off” on the setup screen.
Select “Patient Alarm” to choose between 2 settings: choose
“.5” ” to set the alarm to activate when a 0.5°C difference is
read between a set temperature and the baby probe tempera-
ture, or choose “1.0” to set the alarm to activate when a 1.0°C
difference is read between a set temperature and the baby
probe temperature.”1.0” is the default setting. The “Patient
Alarm” is used in either the incubator or warmer Baby modes.
Select “Canopy Pedal” to disable or enable foot pedals that
raise the canopy. When the pedal is disabled, the canopy can
still be raised or lowered with the buttons on either side at the
head of the bed.
Select “Elevate” to disable or enable the foot pedals that raise
or lower the bed height.
Select “Baby Info” to bring up the Baby Info screen. By turning
the knob until the desired character appears and then pushing
the knob, you can enter up to a 9-character name. Blanks and
periods can also be selected. Selecting “ ” allows you to go
back and change the previous character. Selecting “ ”
enters the letters you have selected for names that are less
than 9 characters long. By selecting the month, day and year
numbers, you can enter the baby’s birth date.
These settings will be saved for 2 hours after power down
Select Done/Exit to return to the option icon screen.
Note: A service screen allows you to change preheat
% power, alarm criteria, disable the canopy control pedal and
other default settings. See the service manual for a description
of this screen and how it may be accessed.
?
Options
Done/Exit
Alarms:
Setup
Temp Units
C
Alarm Volume
Man Temp Alarm
Off
Patient Alarm
1.0C
Canopy Pedal
Enable
Elevate
Enable
Baby Info
Done/Exit
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
CI.31.031
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...