2-2 6600-0342-000 08/08/01
2-2
Preoperative Checkout Procedure
8.
If the unit is equipped with an iris porthole, check that the iris is
installed and in good condition. Check that the iris tightens when
it is rotated.
9.
Check that the tubing access covers in the four bed corners and
the large slot grommet at the head of the bed are in place and are
in good condition.
10. Check the operation of the bed. The bed surface should rotate
easily without binding. If the bed is properly seated and locked in
place, the mattress should be level. When the bed is rotated back
into the straight position, check to see that the bed platform extends
and stops when it is pulled out on either side. Check the operation
of the bed tilt mechanism. When you squeeze the tilt control and
push down on the foot of the bed the head of the bed should raise
easily, and should stay in position at any angle along its tilt path
when you let go of the tilt control. When you push down the head of
the bed the foot of the bed should raise smoothly, and should stay
in position at any angle along its tilt path when you let go of the tilt
control.
Controller checks
WARNING
w
w
w
w
w
Do not use the OmniBed in the presence of
flammable anesthetics: an explosion hazard
exists under these conditions.
1.
Make sure the power cord is connected to the outlet on the unit
and to the rated power supply.
2.
Connect the baby probe to jack 1 on the jack panel at the head of
the bed.
3.
Switch on the power at the mains switch on the back of the unit,
and at the standby switch on the jack panel. Verify the following:
•
All the displays and indicators light
•
The software revision appears
•
The prompt tone begins
Note: If the unit has been used in the last 2 hours, the patient history
query also appears.
4.
Adjust the set temperature to silence the prompt tone.
5.
Check the patient probe. Warm it by placing it between your
fingers, and verify that the baby temperature reading increases.
Note: If the probe temperature is below 30°C, the display will show -
L-. If the unit is in the Baby Mode, there will also be a Baby Cold
alarm.
6.
Unplug the patient probe and check that both visual and audio
alarms trigger in the Baby mode.
7.
Check the canopy lift mechanism. Push the raise canopy pedal and
verify the canopy moves smoothly in one continuous movement to
its upper limit, the heater doors open, and the unit shifts into warmer
operation.
Check that the canopy pedals on both sides of the unit
raise the canopy.
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...