1-12
6600-0342-000 08/08/01
1-12
General Information
Baby Hot - Check Probe 1
Active whenever the unit is in the baby mode, the Baby Hot
alarm activates when the temperature measured by the probe
on the baby skin is 1.0°C above the baby temperature you set
previously (this can be adjusted to 0.5°C using the Setup/
Customize screen). Monitor the patient carefully and check the
attachment of the probe on the baby’s skin. The audio alarm is
suspended for a ten minute period when you push the alarm
silence button. This alarm can also be activated in the Manual
and Air modes. See setup screen.
Baby Cold - Check Probe 1
Active whenever the unit is in the baby mode, the Baby Cold
alarm activates when the temperature measured by the probe
on the baby skin is 1.0°C below the baby temperature you set
previously (this can be adjusted to 0.5°C using the Setup/
Customize screen). Monitor the patient carefully and check the
attachment of the probe on the baby’s skin. The audio alarm is
suspended for a ten minute period when you push the alarm
silence button. This alarm can also be activated in the Manual
and Air modes. See setup screen.
Air Temperature Greater Than... Alarms
These alarms warn of abnormally high temperatures read by
the compartment probe in all control modes and automatically
shut off the heater. They are not self resetting; once the alarms
are activated they must be manually reset by pushing the
alarm silence button, even if the alarm condition is resolved.
Computer independent circuitry continuously measures and
compares the compartment temperature with factory set
levels. Visual and audible alarms are activated and the heater
shuts off if the compartment temperature exceeds these preset
levels, independent of the software.
The functionality of all other alarms is continually checked by
the microprocessor software during normal operation. If a fault
occurs in any of this circuitry, an indicator lights and a tone is
sounded.
Additionally, a computer independent “watchdog” circuit will
reset the computer, turn off the heater, and activate the alarms
in the event of a microprocessor failure or software error.
Air Temperature >38 C
Active only when the canopy is closed and the set temperature
is below 37°C, the Air Temp >38°C alarm activates in the air
mode when temperatures greater than 38°C are read by the
compartment probe. This alarm may occur when you have just
lowered the air set temperature, or when you have changed
from baby to air control mode. Monitor the unit closely to
assure the air temperature decreases.
?
Options
Done/Exit
Baby Hot- Check Probe 1
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
?
Options
Done/Exit
Baby Cold- Check Probe 1
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
?
Options
Done/Exit
Air Temperature > 38 C
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
CI.31.036
CI.31.037
CI.31.038
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...