Performance
System
Control accuracy
± 1.0°C
Control Temp vs. Avg. Incubator Temp with
level bed in manual mode.
Variability
± 0.5°C
Incubator Temp vs. Avg. Incubator Temp
Warm-up time
< 50 min.
Time to reach 39°C control temp from cold
start in 25°C 50% RH room ambient
Patient measurement accuracy ± 0.3°C @ 30°C to 42°C
Accuracy of patient temperature
measurement system within range of
temperature measurement
Air Velocity
< 10 cm/sec
In Whisper Quiet™ mode velocity measured
10 cm above the center of the mattress,
closed bed
CO
2
level
0.2%
Maximum CO
2
level measured per IEC
601-2-19, clause 105.1
Sound level
< 49 dbA
In Whisper Quiet™ mode closed bed sound
level measured 10 cm above the center of
the matress
Humidity
Servo control accuracy
± 10 % for settings up to
Humidity control setting vs. average
85%; minimum 75%
humidity at 10 cm above center of bed
for settings >85%
Manual humidity
>75%
Humidity level maximum setting in 25°C
50% RH room ambient
Ramp-up time
<50 minutes
Time to reach 75% RH with a 39°C control
temp from cold start in 25°C 50% RH room
ambient
Operating time without refill
>12 hours
Operational time at 65% RH control setting
with one filling of reservoir in 25°C 50% RH
room ambient
Weight Scale
Functional range
300 gm to 8 kg
Accuracy
± 10 gm
Resolution
10 gm (factory setting) or 5 gm
Mechanical Specifications
Canopy closed
Canopy closed
Canopy open
Canopy open
bed lowered
bed raised
bed lowered
bed raised
Height:
147 cm
174 cm
208 cm
236 cm
Width:
69 cm
Depth:
112 cm
Weight:
129 kg
Accessories
Maximum load
Storage drawer
7 kg
Monitor shelf
23 kg
Instrument shelf
9 kg
Total each accessory rail
23 kg
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...