General Information
Option Alarms
Humidifier
Add Water- Water level low in humidifier reservoir. Add
distilled water.
Water Reservoir Not Engaged- Humidity water reservoir
missing or incorrectly installed.
Humidity Probe Failure- Probe failed; humidifier not operat-
ing. Check jack connection. Consider increasing air tempera-
ture or removing baby.
Humidity Failure- Heater failed; humidifier not operating.
Repair humidifier.
?
Options
Done/Exit
Add Water
?
Options
Done/Exit
Water Reservoir Not Engaged
?
Options
Done/Exit
Humidity Probe Failure
?
Options
Done/Exit
Humidifier Failure
10:30 am
6/23/00
Name
Birth: 6/21/00
Wed
Scale
Exceeds Maximum Weight- Weight on scale platform is 8
kilograms or more. Check for other objects on scale
Scale failure- System failure; note error code and remove
scale from service.
?
Options
Done/Exit
Exceeds Maximum Weight
?
Options
Done/Exit
Scale Failure 3
CI.31.048
CI.31.049
CI.31.050
CI.31.051
CI.31.052