Maintaining the OmniBed
Maintenance schedule
The unit should be maintained in accordance with the proce-
dures detailed in the Service Manual. Service maintenance
must be performed by a technically competent individual.
Operator maintenance
This schedule lists the minimum frequencies. Always follow
hospital and local regulations for required frequencies.
Weekly or After Each Patient
Disinfect the humidifier if used.
Clean the incubator and check the air filter. Disinfect the
incubator if required or after use with infectious patients.
Quarterly
Replace the air filter (located behind the humidifier reservoir).
When you replace the filter, mark the date on the label sup-
plied with the replacement filter and affix it to the filter cover
panel.
Note: This is the minimum replacement frequency. The filter
must also be replaced whenever it appears dirty, or has been
used with an infectious patient.
Service maintenance
This schedule lists the minimum frequencies. Always follow
hospital and local regulations for required frequencies.
Annually
Perform the electrical safety and calibration procedure as
described in the service manual.
Calibrate the scale as described in the service manual.
Every Two Years
Replace the battery.
Note: The battery is used to sound the power failure alarm and
to power memory circuits during a power failure.
Every Three Years
Calibrate the humidifier as described in the service manual.
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...