![Ohmeda Giraffe OmniBed Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/ohmeda/giraffe-omnibed/giraffe-omnibed_operators-manual_3546691031.webp)
3-3
6600-0342-000 08/08/01
3-3
Operating the OmniBed
5. Check the following:
- All LED segments and Screen pixels light, then go blank
- All installed options appear on graphics screen
- Patient history query appears
- Operator prompt sounds
WARNING
w
The patient probe is not isolated from earth
ground. Any additional equipment used with the
Giraffe OmniBed must comply with IEC 601.
Operation as an Incubator
Always preheat the bed before placing the baby in the bed.
1. Select air control mode. If the baby mode is desired, preheat the
bed in air mode. A suggested thermal neutral temperature can be
calculated using the comfort zone screen.
2. Select a temperature setting using the temperature/power
buttons.
3. Place the baby in the bed.
4. For baby mode operation, connect the skin temperature probe to
probe jack 1. In the air mode, use a probe only if you wish to
display the patient skin temperature. See “Attaching the Skin
Temperature Probe” later in this section.
5. If baby mode is desired, press the Baby mode button.
6. Select a temperature setting using the temperature/power
buttons.
Operation as a Warmer
Always preheat the bed before placing the baby in the bed.
1. Select a control mode (manual mode is the default). Preheat the
bed in the manual mode. Power settings in the preheat range can
be set without causing a check patient alarm. The preheat range is
factory set at 0 to 25%. This range can be changed on the service
screen.
If you start with the canopy open, or open the bed during the first
minute after start up, and you do not select a control mode, or enter
a percentage power setting the graphics screen will show the
“Warm Up Mode” message and the radiant heater will automatically
run at 100% power. When ten minutes has elapsed and you still
have not selected a control mode, the graphics display will show
the “Preheat Zone” message and the radiant heater will then run at
the maximum preheat level.
2. Select power percentage setting using the temperature/power
buttons.
3. Place the baby in the bed.
4. For baby mode operation, connect the skin temperature probe to
probe jack 1. In the manual mode, use a probe only if you wish to
display the patient skin temperature. See “Attaching the Skin
Temperature Probe” later in this section.
5. If baby mode is desired, press the Baby mode button.
6. Select a temperature setting using the temperature/power
buttons.
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...