![Ohmeda Giraffe OmniBed Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/ohmeda/giraffe-omnibed/giraffe-omnibed_operators-manual_3546691038.webp)
3-10
6600-0342-000 08/08/01
3-10
Operating the OmniBed
The Nurse Call alarms works with the OmniBed’s audible alarm.
Silencing the audible alarm on the OmniBed stops the Nurse
Call alarm even if the alarm condition still exists. At the end of
the silence period, the Nurse Call alarm and the audible alarm
reactivate unless the condition has been resolved. The alarm
silence period ends prematurely if another alarm triggers.
Refer to the Appendix for additional information on Nurse Call
connections.
Nurse Call checkout:
1. Complete the basic operating procedure in chapter 3.
2. Verify proper operation of the Nurse Call station.
3. Connect the Nurse Call connector to the OmniBed.
4. Unplug the unit to trigger an alarm. Verify that you also get
an alarm at the Nurse Call station.
Note: Any power interruption (switching off the unit, acci-
dentally unplugging the power cord, etc.) triggers a Nurse
Call alarm.
Using the In-Bed Scale
The OmniBed can be equipped with its own in-bed scale that is
operated from the graphics screen. See accessory options in
chapter 1 for information about the scale menu screens work.
Scale installation
1. Remove the clear plate and mattress from the bed.
2. Fit the scale weighing platform into the recesses in the bed.
3. Route the scale cable though a tubing access cover or the
ventilator slot and plug the connector into the jack panel.
4. Place the clear plate and mattress on the weighing plat-
form. Make sure they are properly seated on the platform.
Weighing procedure
The baby should be in approximately the center of the bed.
Stuffed animals and other objects should not lean against side
walls. All leads, I.V. tubes and ventilator tubes should be
secured. Blankets may be tucked under the mattress, but must
not be tucked under the weighing platform.
1. Make sure the bed tilt platform is completely level; if the
scale is tilted it will effect weighing accuracy.
2. Select the scale icon options display to bring up the scale
menu. If the scale is not connected, no icon will appear.
3. Select weigh from the menu to initiate weighing.
4. In order to “zero” the scale, the display will prompt you to lift
the baby by showing the “Lift Baby” icon while a tone
sounds. Lift the baby and any tubing or leads attached to
the baby. Make sure that arms, legs, blankets, and clothing
are clear of the mattress.
1. Mattress pad
2. Clear plate
3. Scale
4. Bed
1
2
3
4
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...