Maintaining the OmniBed
Disassembling the OmniBed for a
complete cleaning
If the OmniBed has been exposed to body fluids, check your
hospital’s infection control procedures. You may need to wear
protective clothing and eye wear or use special sterilization
agents and cleaning procedures.
1. Raise the canopy to it’s top travel limit.
WARNING
w
w
w
w
w
Disconnect the power cord before
disassembling or reassembling the
lower unit.
CAUTIONS
w
Electronic devices in the microprocessor
controller are susceptible to damage
from discharges of static electricity.
These devices are adequately protected,
but can be damaged if the unit is disas-
sembled beyond that recommended for
cleaning and maintenance.
w
Do not remove the air temperature probe
mounted to the rear wall; the air tem-
perature probe must remain in the
correct position for proper operation.
2. Turn the power switch Off and unplug the unit from the
power outlet.
3. If the OmniBed was previously on, allow it to cool for at
least 30 minutes.
4. Unplug the patient probe and any other probe jacks or
connectors from the jack panel.
5. Open the side doors by squeezing the latches located at
the top corners of each door. To remove the inner wall,
push out on the tabs on the back of the door latches to
release the top of the wall (A), then rotate the inner wall
down until it slides directly out of the door hinges (B).
6. Remove the side doors by pushing in on either of the spring
loaded buttons located at the bottom corners of the door
(C) and lifting the door out of its hinges (D).
7. Lower the wall at the foot of the bed by lifting the wall
directly up, then allowing to rotate forward and down.
8. Because of the compartment probe connection, the wall at
the head of the bed can not be completely removed. Lift the
wall up from the center. The wall will stop approximately 4
cm up to allow cleaning underneath it.
9. Remove the porthole seals and irises.
A
B
C
D
CI.31.081
CI.31.080
CI.31.078
CI.31.079
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...